First beam profiles for PEPITES

**The ultra-thin profiler PEPITES was able to successfully measure its first proton beams at ARRONAX*]

This Tuesday, May 31, 2022 will forever remain a key date for the PEPITES project and all its participants. It was indeed at 10:09 a.m. (!) that the detector, installed the previous week, delivered its first diagnosis in the form of the transverse profiles of the 68 MeV proton beam delivered by the ARRONAX cyclotron. With a dedicated ASIC developed by CEA-DEDIP as part of the project, we were able to “see” beams ranging from 20nA to pA and measured signals as low as 30fA! Great successes for this project started a few years ago and which augurs great work to come.

2D profile of 68 MeV ARRONAX proton beam. Beam current was 1pA.

The PEPITES project aims to build a functional prototype of an ultra-thin monitor of charged particle beams. Developed for the needs of protontherapy, where knowing the parameters of the beams is a crucial irradiation issue, the detector aims to be as less interceptive as possible in order to induce minimal disturbance during the treatment.

In order to maintain a total water-equivalent thickness of the detector below 10 microns, the signal is formed from the production of secondary electrons, a surface phenomenon requiring very little material (about ten nanometers). 50nm thick gold strips are therefore deposited on a substrate, a polyimide film 1.5 microns thick, resistant to radiation. When the incident particles intercept these strips, secondary electrons are emitted and the current thus formed is read on each of the strips, thus making it possible to measure the position and the transverse dimensions of the beam.

The PEPITES detector.

In addition to providing continuous beam monitoring, the low thickness of the monitor allows it to tolerate higher currents than existing systems without suffering from overheating problems and its physical signal formed by the secondary electrons does not saturate. These properties make PEPITES an ideal candidate for measuring future particle beams envisaged for so-called “Flash” radiotherapy, where the entire dose necessary to neutralize a tumor is delivered in less than 100ms. This new treatment modality, still under study, involves very intense beam currents for which PEPITES has undeniable assets allowing their characteristics to be measured.

To know more : https://llrwebpepites.in2p3.fr/